The Charme® reference aerosol electrometer measures the electrical charges on aerosol particles. If the charge status of the aerosols is not known, then a charger or neutralizer is installed upstream from the Charme® in order to generate a defined charge distribution. The charged particles are discharged through very high resistance. The resulting discharging current is measured with a resolution of 1 fA = 10-15 . The Charme® is also used to calibrate CPCs as per ISO 27891 because current measurement is directly traceable to SI units. An on-site correlation between the measured current and the mass concentration is able be determined by means of a gravimetric filter, which can be switched out by the user. The Charme® features reliable performance, components of optimal quality, and easy operation using the touch screen.

Charme®

Charme®

Determination of the electrical charge of ultrafine aerosols from 2 nm to 10 µm for the determination of the number concentration

Newsletter Subscription

Stay up to date with Filter Integrity Ltd

  • This field is for validation purposes and should be left unchanged.
Filter Integrity Ltd | Filtration and Particle Science
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.