The dilution systems from Palas® are characterized unambiguously. This is documented with a calibration certificate for each individual device.
The dilution steps deliver a temporally constant, representative dilution with the factors 10 and 100.
The dilution systems can be cascaded with the factors 100, 1,000, 10,000 and 100,000
Low compressed air consumption , e.g. just 128 l/min with a dilution factor of 10,000 with four VKL 10 systems
The dilution steps are combinable with all common particle counters.
With a simple test set-up these cascaded dilution systems can be checked by the users themselves.
Isobaric dilution up to 10 bar overpressure / isothermal dilution up to 120°C with the VKL 10 E, VKL 10 ED, KHG 10 and KHG 10 D dilution systems
Simple functional test on-site
Description
The VKL 10 series of dilution systems can reduce the concentration of aerosols by the dilution factor 01:10, also of very highly concentrated aerosols, in a defined and reliable way.
The Palas® VKL 10 dilution systems are used in vertical operation for the particle size range up to 20 µm for powders and dusts. Dilution factors of up to 1:100,000 are achieved by cascading several VKL systems.
Functional principle
Particle-free air with the volume flow VR circulates through an annular passage around the suction nozzle. Thus, according to Bernoulli, a volume flow VAn is generated at the suction nozzle.
The dilution factor VF is calculated according to the following formula
Representative dilution of particle size distribution of the Palas® dilution systems by cascading
VDI report no. 1973 from 2007 proved metrologically that a reproducible aerosol dilution is possible with the Palas® dilution systems down to VF 100,000.
Simple functional test on-site
With this simple test set-up, the Palas® cascaded dilution systems can be checked by anyone themselves:
Firstly a particle measurement is performed with one dilution step.
Here it is important that the aerosol concentration, e.g. lab air, to
be measured does not exceed the coincidence limit (maximum detectable
aerosol concentration). In the second step, the dilution step to be
tested is connected in series (cascaded). To check the dilution factor
of the test step (pos. 2), the total particle count form the measurement
in pos. 1 is divided by the total particle count from pos. 2.
Experimental setup
Position 1: Lab air
Position 2: Lab air
The VKL 100 serves to measure coincidence-free with the OPC; the VKL 10 is tested.
Measurement example
Calculation of the dilution factor
Provided the first measurement is not affected by a coincidence
error and the dilution system under test is working (not soiled), a
dilution factor of almost 10 is determined. If this should not be the
case, there was possibly coincidence in measurement 1. In this case the
aerosol concentration has to be decreased or a further dilution step
used. Another possibility would be that the dilution step to be tested
is soiled. In this case the device has to be cleaned and the test
repeated.
Table 1: Technical characteristics of Palas® dilution systems
Application
Aerosol measurement technology: diesel exhaust gases, swarfs, coolant aerosols, weld smoke, oil droplets, test aerosols of filters and inertial separators
Separation efficiency determination with counting measuring methods, e.g. with dust filters or HEPA/ULPA filters
Leak test and acceptance measurements of clean rooms, isolators and safety work benches
Inhalation toxicology
Quality control of respirator masks and filter cartridges
To enquire about the VKL 10 VKL system and the options available contact us now.
This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.