PMFT 1000 M

PMFT 1000 M

PMFT SystemProtective Mask Test System

Testing of respiratory protective face masks.

Exact analysis of filter mask efficiency from 145 nm up to 40 µm.  SARS-CoV-2 size approx. 120 nm – 160 nm.

Benefits
Description
Datasheet
Application

Benefits

 

  • Test rig working principle better than EN 143, EN 149 and EN 13274-7
  • Equivalent to GB 2626, 42 CFR 84 and ASTM 2299-3 by additional software option
  • Test of community masks equivalent to CWA 17553
  • Includes 2 Aerosol generators for NaCl and for oil
  • Testing of fractional efficiency, e.g. efficiency in whole size range of 145 nm up to 40 µm
  • Exact analysis of filter and filter mask efficiency for SARS-CoV-2 (size approx. 120 nm up to 160 nm). Efficiency also displayed at 145 nm
  • Future proof: Works with any kind of aerosol without adjustments
  • Further measurement of differential pressure, e.g. as well within different face velocities to simulate test of breathing resistance
  • Face velocity adjustable between 1.5 – 70 cm/s
  • Product capable of fast quality assurance AND continuous optimization in R&D (display of size distribution)
  • Individual face mask adapter for your product
  • Attractive 2 years maintenance package for availability of test rig
  • Can be operated as Mas Q-Check with optional Mas-Q-Head

The software extension additionally offers:

  • Display of penetration results of the entire tolerance range of the size distribution according to EN 13274-7
  • Allows the comparison of different test institutes and test systems
  • Facilitates certification
  • Shows wide range of standards. Depending on the test operation – i.e. particle size distribution of the test aerosol – one and the same mask can perform very well or fail the test

Description

The PMFT 1000 M tests respirators better than the standard with additional accurate analysis of filter mask efficiency for SARS-CoV-2 (size approximately 120 nm to 160 nm). Indication of efficiency in the virus size range from 145 nm.

The PMFT 1000 M is equipped with aerosol generators for measuring penetration with oil and salt. Measurement procedures for quick quality control (short test) or for testing according to standard (exposure test) are supplied.

Operation and automatic printout of the measurement results are therefore easy even for inexperienced users.

The measurement of total penetration and penetration via particle size is carried out with the high-precision aerosol photometer Promo LED 2300.

The size distribution of the test aerosol according to the standard is as follows:

EN 149:

Oil: Media diameter 290 nm | Geom. standard deviation 1,85
Salt: Media diameter 70 nm | Geom. standard deviation 2.5

42 CFR 84 / GB 2626:

Oil: Media diameter 200 nm | Geom. standard deviation 1.6
Salt: Media diameter 55 nm | Geom. standard deviation 1.86

NIOSH_Verteilungen_Plot.PNG

Based on the measurement of the particle penetration via the particle size, the PMFT can also be used to consider the tolerances in the particle size distribution according to EN 13174-7 as min./max. penetration value. In addition, comparative values to other manufacturers are possible (option).

Comparison of EN 149 penetrations.PNG

The comparison with other standards (42CFR84/NIOSH) is also easy with the 42CFR84 upgrade KIT.

Comparison Niosh.PNG

Overview of standards for face mask penetration testing

EN 149
EN 13274-7
EN 13274-7
GB 2626
GB 2626
42 CFR 84
42 CFR 84
Aerosol
see EN 13274-7
NaCl
PaO
NaCl
PaO/DOP
NaCl
DOP
Mean diameter
see EN 13274-7
0.06 – 0.1 µm
0.29 – 0.45 µm
0.055 – 0.095 µm
0.165 – 0.205 µm
0.055 – 0.095 µm
0.165 – 0.205 µm
Standard deviation
see EN 13274-7
2 – 3
1.6 – 2.2
< 1.86 (by additional software module)
< 1.6 (by additional software module)
< 1.86 (by additional software module)
< 1.6 (by additional software module)
Concentration
see EN 13274-7
4 –12 mg/m³
15 – 25 mg/m³
< 200 mg/m³
(50 mg/m³) < 200 mg/m³
< 200 mg/m³
< 200 mg/m³
Discharge
-
-
-
required
required
required
required
Air flow
see EN 13274-7
95 l/min
95 l/min
85 ± 4 l/min
85 ± 4 l/min
85 ± 4 l/min
85 ± 4.25 l/min
Temperature
see EN 13274-7
22  ± 3 °C
-
25 ± 5 °C
25 ± 5 °C
25 ± 5 °C
25 ± 5 °C
Rel. humidity
see EN 13274-7
< 40 %
-

20 – 40 %

(by compressed air)

-

20 – 40 %

(by compressed air)

20 – 40 %

(by compressed air)

Measurement device
see EN 13274-7

Sodium flame photometer

Light scattering photometer

particle detector

particle detector

Light scattering photometer

Light scattering photometer

Measuring time
see EN 13274-7
30 s
30 s

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

Pause time
see EN 13274-7
180 s
180 s

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

lowest eff. during loading

Exposition
120 mg
120 mg
120 mg

200 ± 5 mg

200 ± 5 mg

200 ± 5 mg

200 ± 5 mg

PMFT remarks

O.K.

O.K.

O.K.

O.K. with upgrade KIT

O.K. with upgrade KIT

O.K. with upgrade KIT

O.K. with upgrade KIT

Application

 

  • Test of respiratory masks
  • Exact analysis of
    filter mask efficiency for e.g. Corona Virus
  • Filter testing for HEPA quality
  • Can be operated as Mas-Q-Check with optional Mas-Q-Head

To enquire about the PMFT 1000 M Protective Mask Test System and the options available contact us now.

"*" indicates required fields

*
This field is for validation purposes and should be left unchanged.

Newsletter Subscription

Stay up to date with Filter Integrity Ltd

  • This field is for validation purposes and should be left unchanged.