PMFT 1000 F

PMFT 1000 F

PMFT System

PMFT 1000 F as an all-rounder in the testing of almost all mask and filter types. Test of total penetration better than the standard, exact analysis of filter mask efficiency from 145 nm to 5 µm

Benefits
Description
Application

Benefits

  • Test rig working principle better than EN 143, ISO 16900-3, 42 CFR 84, EN 149, EN 13274-7 and GB 2626
  • Determination of photometric total penetration for the size range according to standard
  • Includes two aerosol generators for NaCl and for oil
  • Integrated corona discharger CD 2000 A
  • Testing of fractional efficiency, e.g., efficiency in whole size range of 145 nm up to 40 µm
  • Exact analysis of filter and filter mask efficiency for SARS-CoV-2 (size approx. 120 nm up to 160 nm). Efficiency also displayed at 145 nm
  • Future proof: Works with any kind of aerosol without adjustments
  • Further measurement of differential pressure, e.g., as well within different face velocities to simulate test of breathing resistance
  • Face velocity adjustable between 1.5 – 70 cm/s
  • Product for fast quality assurance and continuous optimization in R&D (display of size distribution)
  • Attractive two years maintenance package for availability of test rig
  • Can be operated with standard filter threads according to EN 148-1
  • Measurement of filter efficiencies up to 99.9995 %

Description

The PMFT 1000 F tests respiratory protection particle filters better than the standards EN 143, ISO 16900-3, and NIOSH 42 CFR 84 with additional accurate analysis of filter mask efficiency for SARS-CoV-2 (size approximately 120 nm to 160 nm). In addition, testing face masks based on EN 149/EN 13274-7 standards is possible. Both total photometric penetration and fractional efficiency are tested, e.g., the efficiency in the whole size range, respectively, the particle size-dependent penetration.

Application

  • Testing of full face mask filters for use with standard filter fittings according to EN 148-1
  • Measurement of total penetration of respirators
  • Exact analysis of filter mask efficiency, e.g. coronavirus
  • Media test for HEPA quality

To enquire about the PMFT 1000 F PMFT System and the options available contact us now.

"*" indicates required fields

*
This field is for validation purposes and should be left unchanged.

Newsletter Subscription

Stay up to date with Filter Integrity Ltd

  • This field is for validation purposes and should be left unchanged.
Filter Integrity Ltd | Filtration and Particle Science
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.