A representative sample taking is one of the basic requirements to measure aerosols correctly with a high-quality aerosol spectrometer as e.g the welas® digital or the Promo® system. To have always a repeatable and representative sample taking, Palas® provides the sample taking system ISP 1000.
Important parameters for the sample taking are:
AIPs / Overview sampling probe points AIP for sampling probe volume flow of 5 l/min and standard sampling tube (stainless steel: inner diameter: 6 mm, outer diameter: 8 mm)
Description | For flow velocity in the duct |
AIP 103 | 0.008 – 0.012 m/s |
AIP 86 | 0.012 – 0.017 m/s |
AIP 72 | 0.017 – 0.025 m/s
|
AIP 60 | 0.025 – 0.036 m/s |
AIP 50 | 0.036 – 0.052 m/s
|
AIP 41 | 0.052 – 0.074 m/s
|
AIP 35 | 0.074 – 0.11 m/s
|
AIP 29 | 0.11 – 0.15 m/s
|
AIP 24 | 0.15 – 0.22 m/s
|
AIP 20 | 0.22 – 0.32 m/s
|
AIP 16.6 | 0.32 – 0.46 m/s
|
AIP 13.9 | 0.46 – 0.66 m/s
|
AIP 11.6 | 0.66 – 0.95 m/s
|
AIP 9.6 | 0.95 – 1.37 m/s
|
AIP 8.0 | 1.37 – 1.98 m/s
|
AIP 6.7 | 1.98 – 2.85 m/s
|
AIP 5.6 | 2.8 – 4.1 m/s
|
AIP 4.6 | 4.1 – 5.9 m/s
|
AIP 3.9 | 5.9 – 8.5 m/s
|
AIP 3.2 | 8.5 – 12 m/s
|
AIP 2.7 | 12 – 17.5 m/s
|
AIP 2.3 | 17.5 – 21 m/s
|
AIP 2.0 | 21 – 31 m/s
|